Model 20 and 34 Specifications

LEFTARROW2.GIF (144 bytes)
Back to Products

Model CUJ
COOL Source
Universal Jaw
C
COOL Source
5 Hole Block
HUJ
HEAT Source
Universal Jaw
H
HEAT Source
5 Hole Block
Range °C Ambient to ­30° 30° to 500° 30° to 700°
°F Ambient to ­22° 86° to 932° 86° to 1292°
Range is 50°C (90°F) below ambient
Accuracy1 ±0.1°C (±0.2°F)
for probes 0.25 inch (6.4mm) and smaller

±(0.05% of Setpoint + 0.1°C [0.2°F])
for probes 0.25 inch (6.4mm) and smaller

±0.2°C (0.4°F)
for probes >0.25 to 0.5 inch (6.4 to 12.7mm)

±(0.1% of Setpoint + 0.1°C [0.2°F])
for probes >0.25 to 0.5 inch (6.4 to 12.7mm)

Resolution 0.1°C or °F

0.1°C or °F

0.1° to 537°C (999°F)
1° above
537°C (999°F)1

Stability2

±0.05°C (0.1°F)

±0.05°C (0.1°F)
Repeatability3

±0.05°C (0.1°F)

±0.05°C (0.1°F)
Uniformity X-Y axis
(across flats):

<±0.03°C (0.05°F)

<±0.03°C (0.05°F) <0.1°C (0.2°F)
Z axis (depth):

<0.1°C (0.2°F) from 4 inches
(102mm) deep to 2 inches (51mm) deep

<0.1°C (0.2°F) from 4 inches
(102mm) deep to Jaws
<0.1% of Setpoint
from 4 inches (102mm) deep to 2 inches (51mm) deep
Cool/Heat Time

5 minutes to COOL and Stabilize
from ambient to 0°C (32°F)

10 minutes to HEAT
and stabilize @ 300°C

12 minutes to HEAT
and stabilize @ 600°C

Metal Block Aluminum Aluminum Aluminum Bronze Alloy
Jaw accepts up to
0.5 in. (12.7mm)
Drilled holes 1/8,
3/16, 1/4 inch
(3.2, 4.8, 6.4mm)
4 inches (102mm)
deep.
Drilled holes 1/16, 1/8,
3/16, 1/4, 3/8 inch
(1.6, 3.2, 4.8, 6.4, 9.5mm)
4 inches (102mm) deep.
Other holes available.
Jaw accepts up to
0.5 in. (12.7mm)
Drilled holes 1/8,
3/16, 1/4 inch
(3.2, 4.8, 6.4mm).
4 inches (102mm) deep.
Drilled holes 1/16, 1/8,
3/16, 1/4, 3/8 inch
(1.6, 3.2, 4.8, 6.4, 9.5mm)
4 inches (102mm)
deep. Other holes available.

Cooling Method: Peltier Elements.
1Accuracy for one year with ambient 20°­26°C. Accuracy may be enhanced by using external reference probe.

2Measured for 24 hours with ambient 20°­26°C.
3Measured with Rosemount 162CE as follows: 1. Measure temperature at 100°C. 2. Heat to maximum (500°C for Universal Jaw, 700°C for 5 hole block). 3. Remeasure at 100°C. The difference between 1. and 3. is <0.05°C.

Common Specifications

INTERNAL REFERENCE SENSOR: Pt RTD with repeatability better than 0.05°C after repeated cycling to 700°C.

RAMP MODE:
Settable in 0.1°/min. increments.

AMBIENT TEMPERATURE:
4° to 40°C (39° to 104°F).

CASE DIMENSIONS:
   Model 20:
Aluminum 11.5 in. wide
* 8.4 in. high * 7 in. deep (292 * 213 * 178mm). Removable top 2.5 in. high (64mm).
   Model 34: Aluminum 16.0 in. wide
* 10 in. high * 9 in. deep (406 * 254 * 229mm). Removable top 2.5 in. deep (64mm).
   Power:   90 to 130 VAC
                   210 to 250 VAC (specify option 230)
                   800VA max., 50/60Hz.
 

 

ThermaCal and Universal Jaw are trademarks of ThermaCal Inc. Patents are pending on Universal Jaw.

LEFTARROW2.GIF (144 bytes)
Back to Products


ThermaCal INC.
30275 Bainbridge Road
Solon, OH 44139
Phone:    440/498-1005
Fax:        440/498-1062
E-mail:   thermacal@ameritech.net

Site Designed & Hosted by @industry, Inc.