Model 34 Temperature Source/Measure

LEFTARROW2.GIF (144 bytes)
Back to Products

Model 34

 

The Model 34 Temperature Source/Measure

The Model 34 simplifies the testing of temperature sensors by conveniently housing both the Source and Measure in one package. The Model 34 has four distinct capabilities:

1. A COOL Source (either Universal Jaw or "five hole block) with range down to ­30°C (­22°F)

2. A HEAT Source with range to 500°C (932°F) with Universal Jaw or range to 700°C (1292°F) with a "five hole block." The Model 34   may be ordered with two HEAT Sources and no COOL Source.

3. A Measure for thermocouples, RTDs, and thermistors

4. A Measure for 4­20mA and 20VDC from temperature transmitters

ABOVE:

With the HEAT SOURCE at 100.0°C, the thermocouple under test reads 100.8°C on the MEASURE unit, demonstrating a +0.8°C thermocouple error.

 

Ordering Code COOL COOL HEAT HEAT  

 

Universal Jaw
­30°C (­22°F)
5 Hole Block
­30°C (­22°F)
Universal Jaw
500°C (932°F)
5 Hole Block
700°C (1292°F)
Measure
34-CUJ-HUJ *   *   *
34-HUJ-HUJ     * (2)  

*

34-HUJ-H     * * *
34-C-H   * (see photo)   * *
34-H-H       * (2) *
K, J, T, E, C, S, R, B, and N thermocouple types may be read with 0.1°C or °F resolution.

RTDs (Resistance Temperature Detectors) may be tested in 2-, 3-, or 4-wire modes with alpha values of 0.385 or 0.392. Thermistors (2252 (omega)) are also accommodated.

The outputs of temperature transmitters may be read directly by the MEASURE unit in either mA or DC volts.

MEASURE ACCURACY

RTD: ±(0.05% rdg. + 0.1°C [0.2°F]).

Thermocouple: ±(0.1% rdg. + 0.1°C [0.2°F]).

VDC: ±0.02% rdg.

mA: ±0.1% rdg.

LEFTARROW2.GIF (144 bytes)
Back to Products


ThermaCal INC.
30275 Bainbridge Road
Solon, OH 44139
Phone:    440/498-1005
Fax:        440/498-1062
E-mail:   thermacal@ameritech.net